1.
Radiat Prot Dosimetry ; 148(4): 452-6, 2012 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-21561947

RESUMEN

The aim of this study was to evaluate the image quality of 29 computed tomography (CT) scanners in Brazil and to perform estimations of patient dose and image quality of common CT examinations at these equipment. The volume CT air kerma indexes (C(VOL)) were estimated, using normalised weighted air kerma indexes, supplied by the ImPACT group. The image quality tests were performed using the phantom and accreditation protocol from the American College of Radiology (ACR). The C(VOL) values for head scans varied between 8.7 and 108 mGy. The Hi-res chest examinations presented C(VOL) values varying from 0.4 to 32 mGy. For abdominal scans, the estimated C(VOL) values varied between 4.1 and 94 mGy. This wide variation of air kerma between different centres is related to the scanner type and also to the scanning parameters. The results also showed that the image quality did not attend all ACR CT accreditation requirements.


Asunto(s)
Procesamiento de Imagen Asistido por Computador/métodos , Radiometría/instrumentación , Radiometría/métodos , Tomografía Computarizada por Rayos X/métodos , Abdomen/efectos de la radiación , Aire , Brasil , Diseño de Equipo , Humanos , Fantasmas de Imagen , Radiología/métodos , Tomógrafos Computarizados por Rayos X , Estados Unidos